top of page

For Resistance Thermometers, Thermocouples and Electrical signals

 

Highlights

  • Generation of RDT-, TC- mA- and V signals
  • For mobile on-site calibration

SIKA EC 10 Process Calibrator For Resistance Thermometers, Thermocouples

  • Technical Data
    Resistance thermometers (RTD)
    Generation and measurement of RTD signals: Pt100, Pt200, Pt500, Pt1000, Cu10, Cu50
    Ω-generator function: 0-40 000 Ω
    Ω-measurement function: 0-5500Ω
    Accuracy (of reading + const.): +0.02 %
    Measurement of multi-wire connections: 2/3
    Thermocouples (TC)
    Generation and measurement of TC signals: J, K, T, R, S, B, N, E
    mV-generator function: -100-1100 mV
    mV-measurement function: -50-550 mV
    Accuracy (of reading + const.): ±0.02 %
    Selectable temperature unit: °C/°F
    Internal comparison point ±0.5 °C
    Current (mA)
    Loop current signal generation: 0-22 mA
    Loop current signal measurement: -5-55 mA
    Accuracy (of reading + const.): ±0.02 %
    Current loop supply: 24 V ±10 %, 22 mA
    Voltage (V)
    Voltage signal generation: -1-11 V
    Voltage signal measurement: -5-55 V
    Accuracy (of reading + const.):±0.02 %

    • DAkkS certificate
    • SIKA works calibration certificate
    • Transport case
    • Plug connector 4 mm, red
    • Plug connector 4 mm, black
bottom of page