top of page

This item is a good condition used Fluke 720A Kelvin-Varley Divider. This unit has been performance verified in our lab but not formally calibrated. All pictures are of the actual unit.

 

This item comes with a 6 Month (180 day) warranty. Every pre-owned unit we offer goes through meticulous inspection and NIST calibration to guarantee its reliability and durability.

 

For customers in the USA, all purchases arrive cleared. ALL brokerage or customs fees are prepaid by us.  Your item arrives FOB destination.

 

International Buyers Please Note: Import duties, taxes, and charges are not included in the item price or shipping cost. These charges are the buyer's responsibility. Please check with your country's customs office to determine what these additional costs will be prior to bidding or buying. For more information of clarification, please ask.

 

Will ship worldwide. We make our best efforts to get you the best shipping prices. Contact us for other international rates and we will be pleased to give you a quote. We accept VISA / Mastercard. Paypal is also accepted. Sales to within Canada are subject to 5% GST.  Ontario sales subject to 13% HST.

Fluke 720A Kelvin-Varley Divider Primary Ratio Standard 0.1ppm Linearity

$2,629.00Price
  • Model 720A Kelvin-Varley Divider is a high-resolution primary ratio standard with absolute linearity of 0.1 ppm, temperature coefficient of linearity of 0.1 ppm/ºC, and self-calibration capability.

    • 0.1 ppm resolution, seven decades
    • 0.1 ppm of input absolute linearity
    • Built-in self calibration bridge
    • Front panel self-calibration

    Ratio range

    0 to 1.0 (1.0 input tap) and 0 to 1.1 (1.1 input tap)

    Resolution

    0.1 ppm of input with seven decades

    0.2

    Absolute linearity

    (At calibration temperature and without the use of a correct chart) ±1.0 ppm of input at dial

    settings of 1.1 to 0.1, ±0.1 (10S)1/3 of input at dial settings (S) of 0.1 to 0

    Absolute linearity stability

    (Without self-calibration) ±1.0 ppm of input/yr at dial settings of 1.1 to 0.1, ±1.0 (10S)(2/3) ppm

    of input/yr at dial setting (S) of 0.1 to 0

    Note: Absolute linearity is defined as the linearity between max and min output voltages. The self-calibration procedure may be used at any time to reset absolute linearity to ±0.1 ppm of input.

    Temperature coefficient of linearity

    ±0.05 ppm of input ºC maximum at dial settings of 1.1 to 0.1

    Short-term linearity stability

    Under typical standards laboratory conditions (temperature maintained within ± 1 ºC) and with

    an applied voltage of up to 100V, stability of linearity is ± 0.01 ppm/24 hours and ± 0.1 ppm/30 days

    Power coefficient of linearity

    ±0.2 ppm of input/W max at dial settings of 1.1 to 0.1; ±0.2 (10S)2 ppm of input/W max at dial

    settings (S) of 0.1 to 0

    Maximum end errors

    Zero error at output low: 0.004 ppm of input

    Zero error at input low: 0.05 ppm of input

    Full-scale error: 0.05 ppm of input

    Maximum input voltage

    1000 V on 1.0 input terminal, 1100 V on 1.1 input terminal

    Thermal voltages

    ±0.5 μV max

    Input resistance

    100 kΩ ±0.005% at 1.0 input terminal at 25 °C, 110 kΩ ±0.005% at 1.1 input terminal at 25 °C

    Temperature coefficient of input

    resistance

    ±1 ppm per °C max

bottom of page